Zhangjiagang Emergency Medical Service Technology Co., Ltd.
Products
Contact Us
  • Contact Person : Mr. Zhao Alex
  • Company Name : Zhangjiagang Emergency Medical Service Technology Co., Ltd.
  • Tel : 0086-512-58198789
  • Fax : 0086-512-56930918
  • Address : Jiangsu,zhangjiagang,zhangjiagang City, Jiangsu province
  • Country/Region : China
  • Zip : 215600

EMS-100 (Standard) Infant Radiant Warmer

EMS-100 (Standard) Infant Radiant Warmer
Product Detailed
infant radiant warmer 1.Tilt angle of infant bed is adjustable 2.Transparent protector folded 3.Temperature deviation

                                               EMS-100   (Standard) Infant Radiant Warmer

Brief introduction:

1. Microprocessor based servo controlled temperature system.

2. The microprocessor-controlled unit allows for three settings: Pre-warm, manual or automatic.

3. Alarm functions for safety includes: Power failure, temperature deviation, Temperature sensor failure, over temperature.

4. APGAR timer to record therapy time

5. Convenient X-ray cassette under infant bed

6. Heater head can be adjusted +90 degree horizontally

7. Tilt angle of infant bed is adjustable

8. Observe light angle is adjustable

9. Transparent protector folded

10. Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source, the heating is quick and life-span is long

11. Alarms are audible and visual

12. A second thermal cut-out function for more safety

13. Skin temperature sensor failure protect function avoid over temperature

14. Temperature deviation can be amended in front panel directly

Standard configuration

Main body (including the radiant source, control system, infant bed, bracket), X-ray cassette, Observe light, I. V. Pole, skin temperature sensor, tray, mattress, transparent protector, castors.

Optional parts: Drawer, RS-232(485) connecter

Technical data:

1. Power supply: 110/220V 60/50Hz

2. Power input: ≤ 1200VA

3. Operating condition

An ambient Temperature: 18°C-30°C

An ambient Relative Humidity: 30--75 percent RH

Atmospheric Pressure: 700~1060 hPa

An ambient Air Velocity <0.3m/s

4. Temperature control: 32~38°C

5. Temperature control accuracy: ≤ 0.5°C

6. Mattress temperature uniformity: <25°C

7. Skin temperature sensor accuracy: +/- 0.3°C

8. Warm-up time: ≤ 30min

9. Infant bed tilt angle: + 10degree;

10. APGAR timer range: ≤ 24h

11. Heater head horizontal: 90degree;

12. Mattress size: 67cm X 54cm

13. Distance from heater to mattress: 80cm

Transport and storage

Environment Temperature: -40~ +55°C

Environment Relative Humidity: ≤ 95percert Atmospheric Pressure: 500~1060hPa

Package:

Each unit is packed one case

Size: 120cmX 70cm X 96cm

Gross weight: ≤ 90kg

EMS-100 (Standard) Infant Radiant Warmer